MFT#

Quality summary#

QualitySummaryGood QualitySummaryBad

The left panel summarises all MFT QC checks for the last QC cycle. The top line gives aggregated quality status with the text message suggesting actions for the QC shifter. * Quality: BAD: contact MFT on-call expert * Quality: Medium: create a log entry * Quality: NULL: the plots are empty.

The bottom line shows an example of BAD status in Chips in error, Digit occupancy and Cluster occupancy plots.

The right panel provides a time trend of the MFT QC summary. In cosmics, it usually takes ~5 minutes after the start of run for the occupancy plots to be files. If the quality stays BAD after ~5 minutes from the start of run, call the MFT on-call.

If MFT quality is BAD for the whole run duration, MFT should be tagged as BAD in the logbook.


Chips in Error/Fault/Warning#

[QC on FLP - plot reset at each cycle] ReadoutChipsError ReadoutChipsFault ReadoutChipsWarning

Plot description

This plot is created on the FLPs. The histogram shows the number of MFT chips in Error/Fault/Warning and provides their list up to 20 entries.

Checks to be done

Check the message on the histogram.

Actions to be taken

Follow the instructions on the histogram message.


Digit Occupancy Summary#

[QC on FLP - plot integrated over the run]DigitOccSummary

Plot description

This plot is created on the FLPs. The histogram shows the number of digits per ROF (Readout Frame) per MFT zone.

Checks to be done

Wait for ~5 minutes after the start of run for the histogram to be filled. Then check the message in the automatic checker for empty ladders.

Actions to be taken

If two adjacent MFT zones are empty after ~5 minutes from the start of run, the quality is bad (message in red), call the MFT on-call immediately.


Cluster Occupancy Summary#

[QC on EPN - plot integrated over the run]ClusterOccSummary

Plot description

This plot is created on the EPNs. The histogram shows the number of clusters per MFT zone.

Checks to be done

Wait for ~5 minutes after the start of run for the histogram to be filled. Then check the message in the automatic checker for empty ladders.

Actions to be taken

If two adjacent MFT ladders are empty after ~5 minutes from the start of run, the quality is bad (message in red), call the MFT on-call immediately.


Distribution of the #clusters per ROF#

[QC on EPN - plot integrated over the run]ClustersROF

Plot description

This plot is created on the EPNs. The histogram shows the ROF size for clusters.

Checks to be done

The histogram should look similar to the good quality reference in this documentation.

Actions to be taken

If the plot looks significantly different from the good quality reference, call the MFT on-call.


Known issues#