If some of the DDL presents entries that are larger than the others, a red message will appears: please call EMCAL oncall.
If the plot is empty, and EMCAL is included in the data taking, call EMCAL oncall.
[QC on FLP - plot integrated over the run] Bunch minimum amplitude EMCAL+DCAL#
good quality
bad quality
empty
One peak should be visibile if EMCAL is in data taking.
If a second peak appears around "Min raw amplitude (ADC)" in the range 20-50, call EMCAL oncall
If the plot is empty, and EMCAL is included in the data taking: please call EMCAL oncall
[QC on EPN - plot integrated over the run] Cell Occupancy plots (PHYS) for E>0.2 GeV and E<0.2 GeV and Cell Occupancy plots (CALIB) for E>0.5 GeV#
good quality
bad quality (superMod not sending data)
bad quaity (superMod noisy)
Cell Occ. with E>0.2 GeV: the occupancy plots should be uniformly filled during Cosmics runs.
- if something noisy appears, please call the EMCAL oncall.
- If the plot is empty, and EMCAL is included in the data taking, call EMCAL oncall.
Cell Occupancy plots with E>0.5 GeV (CAL)
- if some large empty areas (expect for the PHOS hole, present in the example for Cell Occ. with E<0.2 GeV) appear, please call EMCAL oncall.
[QC on EPN - plot integrated over the run] Cell Amplitude#
no instructions for the moment.
If the plot is empty, and EMCAL is included in the data taking, call EMCAL oncall.
If peaks appear but the text in histogram is okay, That's a problem due to bad channel. send a message to EMCAL expert via mattermost.
[QC on EPN - plot integrated over the run] Cell Time#
If the plot is empty, and EMCAL is included in the data taking, call EMCAL oncall.
If the errors appear during nigh shift:
add a note in the logbook, don't call the on-call.
in case LHC is preparing for collisions the on-call should be notified at "Prepare Ramp" (even during the night in this case) in case of such errors occur in a technical run before, so that the cards can be recovered before taking physics data.