Skip to content

ZDC#

Baseline [QC on FLP]#

| Green panel: good quality | |:--------------------------:| baseline_good

The plot shows the baseline mean values of each ZDC channel. The quality of the histogram is determined by verifying the deviation from the expected average value. If the plot is empty or the messagge start with Warning or Error call the on-call.

Align Plot [QC on FLP]#

Align_good

The plot shows if all channels are aligned. Most of the channels should be centered on sample 6. If any channel deviates by plus or minus one, it is considered good. Otherwise call the on-call

Bc Overflow [QC on FLP]#

Bc_good

The Bc Overflow plot must be empty. If it fills up it means that there are corrupted data. Call ZDC on-call

Bc Data Loss [QC on FLP]#

Data_loss_good

The Data_loss plot must be empty. If it fills up it means that there are problem about data loss. Call ZDC on-call

Known issues#

FLP InfoLoger: message: "Baseline Error in PED_Z%" message: "Rec Error in ADC_Z%" Detector ZDC. During SYNTHETIC runs this can be ignored.